FLUKE Hart 7007 6054 6055高穩定度溫度校正槽 Deep-Well Baths

高穩定度溫度校正槽 Model 7007 6054 6055
Deep-Well Baths

特點:
1.採用矽油、酒精、水 …… 等
2.高穩定度 0.015 度 C 到 0.0008 度 C
3.高均勻性 0.025 度 C 到 0.002 度 C
4.溫度產生採用三明治原理(加熱、等溫、冷卻)一體成形






Specifications 
7007 
6054 
6055 
Range 
–5 °C to 110 °C 
50 °C to 300 °C 
200 °C to 550 °C 
Stability 
±0.001 °C at 0 °C
 (ethanol)
±0.003 °C at 100 °C
(oil 5012) 
±0.003 °C at 100 °C
(oil 5012)
±0.005 °C at 300 °C
(oil 5017) 
±0.003 °C at 200 °C
(salt)
±0.01 °C at 550 °C
(salt) 
Uniformity 
±0.004 °C at 0 °C
(ethanol)
±0.007 °C at 100 °C
(oil 5012) 
±0.007 °C at 100 °C (oil 5012)
±0.015 °C at 300 °C (oil 5017) 
±0.005 °C at 200 °C
(salt)
±0.010 °C at 550 °C
(salt) 
Temperature Setting 
Digital display with push-button data entry 
Set-Point Resolution 
0.002 °C, high res. 0.00003 °C 
0.01 °C, high res. 0.00018 °C 
Display Temperature Resolution 
0.01 °C 
Digital Setting Accuracy 
±1 °C 
Digital Setting Repeatability 
±0.005 °C 
±0.01 °C 
Heaters 
250 to 1000 W 
250 to 1000 W 
225 to 1800 W 
Working Area 
178 mm dia. (7 in) 
196 mm dia. (7.7 in) 
107 mm dia. (4.2 in) 
Depth 
610 mm (24 in) deep, 178 mm dia.
(7 in), removable polycarbonate cover 
610 mm deep (24 in), 196 mm dia. (7.7 in), removable SST lid 
432 mm deep (17 in), 107 mm dia. (4.2 in), removable SST lid, special viewing channel for LIG sighting 
Wetted Parts 
304 stainless steel 
Power 
230 VAC (±10 %), 50 or 60 Hz, 14 A
(Specify frequency, contact Hart if CE mark required.) 
230 VAC (±10 %), 50/60 Hz, 10.7 A 
230 VAC (±10 %), 50/60 Hz, 7.8 A 
Volume 
42 liters (11.2 gal.) 
50 liters (13.2 gal.) 
19.8 liters
(5.2 gal., 43 kg [95 lb.] of bath salt) 
Size (DxWxH) 
470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in),
1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel 
572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel 
572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel 
Distance from Line of Sight to Top of Fluid 
9.5 mm (3/8 in) 
15.9 mm (5/8 in) 
Automation Package 
Interface-it software and RS-232 computer interface are included for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE. 

Fluke Hart 7008.7040.7037.7012.7011 高穩定度溫度校正槽


Fluke Hart 高穩定度溫度校正槽 Model 7008.7040.7037.7012.7011
Fluke Hart Refrigerated temperature calibration baths
半導體IC環溫測試的最愛機種
搭配矽油可置放多個IC或IC板

溫度校正用的標準溫度源。
1.採用矽油、酒精、水 …… 等
2.高穩定度 0.015 度 C 到 0.0008 度 C
3.高均勻性 0.025 度 C 到 0.002 度 C




Specifications 
7008 
7040 
7037 
7012 
7011 
Range 
–5 °C to 110 °C 
–40 °C to 110 °C 
–10 °C to 110 °C 
Stability 
±0.0007 °C at 25 °C (water)
±0.001 °C at 25 °C
(mineral oil) 
±0.002 °C at –40 °C (ethanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012) 
±0.0008 °C at 0 °C (ethanol)
±0.0008 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012) 
Uniformity 
±0.003 °C at 25 °C (water)
±0.004 °C at 25 °C
(mineral oil) 
±0.004 °C at –40 °C (ethanol)
±0.002 °C at 25 °C (water)
±0.004 °C at 100 °C (oil 5012) 
±0.003 °C at 0 °C (ethanol)
±0.002 °C at 25 °C (water)
±0.004 °C at 100 °C (oil 5012) 
Temperature Setting 
Digital display with push-button data entry 
Set-Point Resolution 
0.002 °C; high-resolution mode, 0.00003 °C 
0.01 °C; high-resolution mode, 0.00007 °C 
0.002 °C; high-resolution mode, 0.00003 °C 
Display Resolution 
0.01 °C 
Digital Setting Accuracy 
±1 °C 
Digital Setting Repeatability 
±0.01 °C 
±0.005 °C 
Heaters 
500 and 1000 Watts 
Access Opening
(call for customs) 
324 x 184 mm
(12.75 x 7.25 in) 
127 x 254 mm
(5 x 10 in) 
162 x 292 mm
(6.38 x 11.5 in) 
127 x 254 mm
(5 x 10 in) 
Depth 
331 mm (13 in) 
305 mm (12 in) 
457 mm (18 in) 
305 mm (12 in) 
Wetted Parts 
304 stainless steel 
Power 
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC, 50 or 60 Hz, 8 A, specify 
115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 or 60 Hz, 9 A
(specify voltage and frequency) 
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify 
Volume 
42 liters (11.2 gal) 
27 liters (7.2 gal) 
42 liters (11.2 gal) 
27 liters (7.2 gal) 
Weight 
61 kg (135 lb) 
63.5 kg (40 lb) 
68 kg (150 lb) 
56.7 kg (125 lb) 
Size (HxWxD) 
610 x 775 x 483 mm (24 x 30.5 x 19 in) 
622 x 768 x 483 mm (24.5 x 30.25 x 19 in) 
775 x 768 x 483 mm (30.5 x 30.25 x 19 in) 
762 x 686 x 401 mm (30 x 27 x 15.8 in) 
559 x 686 x 401 mm (22 x 27 x 15.8 in) 
Automation Package 
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. (Interfaces not available for Model 7011.) 

Fluke Hart 7015, 7009, 7008 高穩定度溫度校正槽

Fluke Hart 高穩定度溫度校正槽 Model7015, 7009, 7008
Fluke Hart Resistor Baths

溫度校正用的標準溫度源。
1.溫度產生採用三明治原理(加熱、等溫、冷卻)一體成形
2.可預先設定 8 種不同溫度在控制面板上
3.如有需要可加裝 RS-232 及 IEEE-488
4.溫度範圍從 -100 度 C 到 +550 度 C






Specifications 
7015 
7009 
7108 
Range 
0 °C to 50 °C 
0 °C to 50 °C 
20 °C to 30 °C 
Stability at 25 °C 
±0.0007 °C  (water)
±0.001 °C  (mineral oil 5011) 
±0.002 °C (water)
±0.004 °C (mineral oil 5011) 
Uniformity 
±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011) 
±0.005 °C (water)
±0.008 °C (mineral oil 5011) 
Temperature Setting 
Digital display with push-button data entry 
Set-Point Resolution 
0.002 °C; high-resolution mode, 0.00003 °C 
Display Resolution 
0.01 °C 
Digital Setting Accuracy 
±1 °C 
±0.5 °C 
Digital Setting Repeatability 
±0.01 °C 
Heaters 
500 and 1000 Watts 
Peltier heating/cooling 
Cooling Capacity 
100 to 200 Watts 
100 W in ambient 23 °C 
Access Opening 
699 x 279 mm (27.5 x 11 in) 
699 x 559 mm (27.5 x 22 in) 
356 x 356 mm (14 x 14 in) 
Bath Chamber Dimensions (HxWxD) (unobstructed space) 
699 x 279 x 330 mm
(27.5 x 11 x 13 in) 
 669 x 559 x 330 mm
(27.5 x 22 x 13 in) 
356 x 203 x 355 mm
(14 x 8 x 14 in) 
Depth 
330 mm (13 in) 
203 mm (8 in) 
Wetted Parts 
304 stainless steel 
Tank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum 
Safety Cutout 
Factory-set high temperature 
n/a 
Power 
115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify 
230 VAC (±10 %), 50 or 60 Hz, 12 A
(specify frequency) 
115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify 
Volume 
95 liters (25 gallons) 
167 liters (44 gallons) 
51 liters (13.2 gallons) 
Weight 
141 kg (310 lb.) 
150 kg (330 lb.) 
35 kg (75 lb.) 
Size (HxWxD) 
1219 x 1118 x 559 mm
(48 x 44 x 22 in) 
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in) 
489 x 413 x 559 mm
(19.25 x 22 x 25 in) 
Automation Package 
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. 
Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50 °C to protect standard resistors placed in the bath.

Fluke Hart 7060.7080 高穩定度溫度校正槽

Fluke Hart 高穩定度溫度校正槽 Model 7060.7080
Fluke Hart Really Cold Temperature Calibration Baths

溫度校正用的標準溫度源
1.採用矽油、酒精、水 …… 等
2.高穩定度 0.015 度 C 到 0.0008 度 C
3.高均勻性 0.025 度 C 到 0.002 度 C








Specifications
70607080
Range−60 °C to 110 °C−80 °C to 110 °C
Stability±0.0025 °C at −60 °C (methanol)
±0.002 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
±0.0025 °C at −80 °C (methanol)
±0.0015 °C at 0 °C (methanol)
±0.0015 °C at 25 °C (water)
±0.003 °C at 100 °C (oil 5012)
Uniformity±0.005 °C at −60 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
±0.007 °C at −80 °C (methanol)
±0.005 °C at 0 °C (methanol)
±0.003 °C at 25 °C (water)
±0.005 °C at 100 °C (oil 5012)
Temperature SettingDigital display with push-button data entry
Set-Point Resolution0.01 °C; high-resolution mode, 0.00007 °C
Display Resolution0.01 °C
Digital Setting Accuracy±1 °C
 
Digital Setting Repeatability±0.01 °C
 
Heaters500 and 1000 Watts
Access Opening127 x 254 mm (5 x 10 in)
Depth305 mm (12 in)
Wetted Parts304 stainless steel
Power230 VAC (±10 %), 50 or 60 Hz, 13 A, single phase, specify frequency
Volume27 liters (7.2 gallons)
Weight159 kg (350 lb.)
Size (HxWxD)
 
1168 x 775 x 483 mm (46 x 30.5 x 19 in)
 
Automation Package

 
Interface-it software and an RS-232 computer interface is included for setting the bath temperature via an external computer. For IEEE-488, add 2001-IEEE.

Fluke Hart 6330.7320.7340.7380 高穩定度溫度校正槽

Fluke Hart 高穩定度溫度校正槽 Model 6330.7320.7340.7380
Fluke Hart Compact Constant Temperature Calibration Baths

溫度校正用的標準溫度源。








Specifications
.
6330
7320
7340
7380
Range
35 °C to 300 °C
–20 °C to 150 °C
–40 °C to 150 °C
–80 °C to 100 °C
Stability
± 0.005 °C at
100 °C
(oil 5012)
± 0.010 °C
at 200 °C (oil 5017)
± 0.015 °C at
300 °C
(oil 5017)
± 0.005 °C at
–20 °C (ethanol)
± 0.005 °C at
25 °C (water)
± 0.007 °C at
150 °C (oil 5012)
± 0.005 °C at
–40 °C (ethanol)
± 0.005 °C at
25 °C (water)
± 0.007 °C at
150 °C (oil 5012)
± 0.006 °C at
–80 °C (ethanol)
± 0.010 °C at
0 °C (ethanol)
± 0.010 °C at
100 °C (oil 5012)
Uniformity
± 0.007 °C at
100 °C
(oil 5012)
± 0.015 °C
at 200 °C (oil 5017)
± 0.020 °C at
300 °C
(oil 5017)
± 0.005 °C at
–20 °C (ethanol)
±0.005 °C at
25 °C (water)
±0.010 °C at
150 °C
(oil 5012)
±0.006 °C at
–40 °C (ethanol)
±0.005 °C at
2 5 °C (water)
±0.010 °C at
150 °C (oil 5012)
±0.008 °C at
–80 °C (ethanol)
±0.012 °C
at 0 °C (ethanol)
±0.012 °C at
100 °C
(oil 5012)
Heating Time
250 minutes,
from 35 °C
to 300 °C
(oil 5017)
80 minutes,
from 25 °C
to 150 °C
(oil 5012)
60 minutes,
from 25°C
to 150°C
(oil 5012)
25 minutes,
from 25 °C
to 100 °C
(oil 5010)
Cooling Time
n/a
100 minutes, from 25°C to –20°C
(oil 5012)
110 minutes,
from 25°C
to –40°C
(ethanol)
130 minutes,
from 25 °C
to –80 °C
(ethanol)
Stabilization Time
15–20 minutes
Temperature Setting
Digital display with push-button data entry
Set-Point Resolution
0.01°; 0.00018° in high-resolution mode
0.01°
Display Resolution
0.01 °
Digital Setting Accuracy
± 0.5 °C
Digital Setting Repeatability
± 0.01 °C
Access Opening
94 x 172 mm (3.7 x 6.8 in)
86 x 114 mm (3.25 x 4.5 in)
Working Area
81 x 133 mm (3.2 x 5.25 in)
86 x 114 mm (3 x 4 in)
Depth
234 mm (9.25 in)
178 mm (7 in)
Wetted Parts
304 stainless steel
Power
115 V ac (±10 %), 50/60 Hz, 7 A or 230 V ac (±10 %), 50/60 Hz, 3.5 A, specify
115 V ac (±10 %), 60 Hz, 15 A or 230 V ac
(±10 %), 50 Hz, 8 A, specify, 1400 VA
115 V ac (±10 %) 60 Hz, 16 A or 230 V ac
(±10 %),50 Hz, 8 A, specify
Volume
9.2 liters (2.4 gal)
4 liters (1 gal)
Size (WxDxH)
305 x 546 x 470 mm (12 x 21.5 x 18.5 in)
off cart; 305 x 546 x 819 mm (12 x 21.5 x 32.25 in) on cart
305 x 622 x 584 mm (12 x 24.5 x 23 in) off cart; 305 x 622 x 819 mm (12 x 24.5 x 32.25 in) on cart
305 x 610 x 762 mm (12 x 24 x 30 in)
Weight
19 kg (42 lb)
35.4 kg (78 lb)
52 kg (115 lb)
Automation PackageInterface-it software and RS-232 included (IEEE-488 optional)
†Rated at nominal 115 V (or optional 230 V)

Fluke Hart 6331.7321.7341.7381 高穩定度溫度校正槽

Fluke Hart 高穩定度溫度校正槽 Model 6331.7321.7341.7381
Fluke Hart Deep-Well Compact Temperature Calibration Baths

1.可預先設定 8 種不同溫度在控制面板上
2.如有需要可加裝 RS-232 及 IEEE-488
3.溫度範圍從 -80度 C 到 +300度 C






Specifications 
6331 
7321 
7341 
7381 
Range 
35 °C to 300 °C 
–20 °C to 150 °C 
–45 °C to 150 °C 
–80 °C to 110 °C 
Stability 
± 0.007 °C at 100 °C
(oil 5012)  ± 0.010 °C at 200 °C
(oil 5017)  ± 0.015 °C at 300 °C
(oil 5017) 
± 0.005°C at –20°C (ethanol)  ± 0.005°C at 25°C (water)  ± 0.007°C at 150°C (oil 5012) 
± 0.005°C at –45°C (ethanol)  ± 0.005°C at 25°C (water)  ± 0.007°C at 150°C (oil 5012) 
± 0.006°C at –80°C (ethanol)  ± 0.005°C at 0°C (ethanol)  ± 0.005°C at 100°C (oil 5012) 
Uniformity 
± 0.007 °C at 100 °C
(oil 5012)  ± 0.017 °C at 200 °C
(oil 5017)  ± 0.025 °C at 300 °C
(oil 5017) 
± 0.007 °C at –20 °C (ethanol)  ± 0.00 7 °C at 2 5°C (water)  ± 0.010°C at 150 °C (oil 5012) 
± 0.007 °C at –45 °C (ethanol)  ± 0.007 °C at 25 °C (water)  ± 0.010 °C at 150 °C
(oil 5012) 
± 0.007 °C at –80 °C (ethanol)  ± 0.007 °C at 0 °C (ethanol)  ± 0.007 °C at 100 °C
(oil 5012) 
Heating Time† 
130 minutes, from 40 °C to 300 °C  (oil 5017) 
120 minutes, from 25 °C to 150 °C  (oil 5012) 
120 minutes, from 25 °C to 150 °C (oil 5012) 
60 minutes, from 25 °C to 100 °C (oil 5012) 
Cooling Time† 
14 hours, from 300 °C to 100 °C (oil 5017) 
110 minutes, from 25 °C to –20 °C (ethanol) 
130 minutes, from 25°C to –45°C (ethanol) 
210 minutes, from 25 °C to
–80 °C (ethanol) 
Stabilization Time 
15–20 minutes 
Temperature Setting 
Digital display with push-button data entry 
Set-Point Resolution 
0.01 °; 0.00018 ° in high-resolution mode 
Display Resolution 
0.01 ° 
Digital Setting Accuracy 
± 1 °C 
Digital Setting Repeatability 
± 0.01 °C 
Access Opening 
120 x 172 mm (4.7 x 6.8 in) 
Depth 
457 mm (18 in) without Liquid-in-Glass Thermometer Cal Kit  482 mm (19 in) with Liquid-in-Glass Thermometer Cal Kit 
Wetted Parts 
304 stainless steel 
Power 
115 V ac (± 10 %), 50/60 Hz, 14.8 A or 230 V ac (± 10 %), 50/60 Hz, 7.4 A, specify 
115 V ac (± 10 %), 60 Hz, 14 A or 230 V ac (± 10 %), 50 Hz, 7 A, specify 
115 V ac (± 10 %), 60 Hz, 16 A or 230 V ac (± 10 %), 50 Hz, 8 A, specify 
230 V ac (± 10 %), 50 or 60 Hz, specify, 10 A 
Volume 
15.9 liters (4.2 gal) 
Size (HxWxD) 
1067 x 356 x 788 mm (940 mm from floor to tank access opening)
[42 x 14 x 31 in  (37 in from floor to tank access opening)] 
Weight 
41 kg (90 lb) 
62 kg (137 lb) 
68 kg (150 lb) 
91 kg (200 lb) 
Automation Package 
Interface-it software and RS-232 included (IEEE-488 optional) 
Rated at nominal 115 V (or optional 230 V)

Fluke Hart 6102 7102 7103 微型校正槽

Fluke Hart微型校正槽 Model 6102 7102 7103
Micro-Baths cc

用於現場溫度校正用的溫度標準源
1.重量輕攜帶方便
2.溫度範圍從-30 度 C 到+200 度 C
3.穩定度從± 0.015 到 0.05 度 C
4.均勻性± 0.02 度







規格:

 
6102 
7102 
7103 
Range 
35 °C to 200 °C (95 °F to 392 °F) 
–5 °C to 125 °C (23 °F to 257 °F) 
–30 °C to 125 °C (–22 °F to 257 °F) 
Accuracy 
± 0.25 °C 
Stability 
± 0.02 °C at 100 °C (oil 5013)
± 0.03 °C at 200 °C (oil 5013) 
± 0.015 °C at –5 °C (oil 5010)
± 0.03 °C at 121 °C (oil 5010) 
± 0.03 °C at –25 °C (oil 5010)
± 0.05 °C at 125 °C (oil 5010) 
Uniformity 
± 0.02 °C 
Resolution 
0.01 °C/°F 
Operating Temperature 
5 °C to 45 °C 
Heating Time 
25 °C to 200 °C: 40 minutes 
25 °C to 100 °C: 30 minutes 
25 °C to 100 °C: 35 minutes 
Cooling Time 
200 °C to 100 °C: 35 minutes 
25 °C to 0 °C: 30 minutes 
25 °C to –25 °C: 45 minutes 
Well Size 
64 mm dia. x 140 mm deep (2.5 x 5.5 in) (working area is 48 mm [1.9 in] in diameter) 
Size (WxHxD) 
14 x 26 x 20 cm
(5.5 x 10.38 x 8 in) 
18 x 31 x 24 cm
(7.2 x 12 x 9.5 in) 
23 x 34 x 26 cm
(9 x 13.2 x 10.5 in) 
Weight 
4.5 kg (10 lb) with fluid 
6.8 kg (15 lb) with fluid 
9.8 kg (22 lb) with fluid 
Volume 
0.75 L (1.6 pints) 
0.75 L (1.6 pints) 
1.0 L (2.11 pints) 
Power 
115 V ac (± 10 %), 2.3 A or 230 V ac (± 10 %), 1.1 A, switchable, 50/60 Hz, 270 W 
115 V ac (± 10 %), 1.8 A or 230 V ac (± 10 %), 0.9 A, switchable, 50/60 Hz, 200 W 
94–234 V ac (± 10 %), 50/60 Hz, 400 W 
Computer Interface 
RS-232 included with free Interface-it software 
NIST-Traceable Calibration 
Data at 50 °C, 100 °C, 150 °C, and 200 °C 
Data at –5 °C, 25 °C, 55 °C, 90 °C, and 121 °C 
Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, and 125 °C