Fluke Hart 高穩定度溫度校正槽 Model 7008.7040.7037.7012.7011
Fluke Hart Refrigerated temperature calibration baths半導體IC環溫測試的最愛機種
搭配矽油可置放多個IC或IC板
溫度校正用的標準溫度源。
1.採用矽油、酒精、水 …… 等
2.高穩定度 0.015 度 C 到 0.0008 度 C
3.高均勻性 0.025 度 C 到 0.002 度 C
Specifications |
7008
|
7040
|
7037
|
7012
|
7011
|
Range |
–5 °C to 110 °C
|
–40 °C to 110 °C
|
–10 °C to 110 °C
| ||
Stability |
±0.0007 °C at 25 °C (water)
±0.001 °C at 25 °C (mineral oil) |
±0.002 °C at –40 °C (ethanol)
±0.0015 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) |
±0.0008 °C at 0 °C (ethanol)
±0.0008 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) | ||
Uniformity |
±0.003 °C at 25 °C (water)
±0.004 °C at 25 °C (mineral oil) |
±0.004 °C at –40 °C (ethanol)
±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) |
±0.003 °C at 0 °C (ethanol)
±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) | ||
Temperature Setting |
Digital display with push-button data entry
| ||||
Set-Point Resolution |
0.002 °C; high-resolution mode, 0.00003 °C
|
0.01 °C; high-resolution mode, 0.00007 °C
|
0.002 °C; high-resolution mode, 0.00003 °C
| ||
Display Resolution |
0.01 °C
| ||||
Digital Setting Accuracy |
±1 °C
| ||||
Digital Setting Repeatability |
±0.01 °C
|
±0.005 °C
| |||
Heaters |
500 and 1000 Watts
| ||||
Access Opening (call for customs) |
324 x 184 mm
(12.75 x 7.25 in) |
127 x 254 mm
(5 x 10 in) |
162 x 292 mm
(6.38 x 11.5 in) |
127 x 254 mm
(5 x 10 in) | |
Depth |
331 mm (13 in)
|
305 mm (12 in)
|
457 mm (18 in)
|
305 mm (12 in)
| |
Wetted Parts |
304 stainless steel
| ||||
Power |
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC, 50 or 60 Hz, 8 A, specify
|
115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 or 60 Hz, 9 A
(specify voltage and frequency) |
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify
| ||
Volume |
42 liters (11.2 gal)
|
27 liters (7.2 gal)
|
42 liters (11.2 gal)
|
27 liters (7.2 gal)
| |
Weight |
61 kg (135 lb)
|
63.5 kg (40 lb)
|
68 kg (150 lb)
|
56.7 kg (125 lb)
| |
Size (HxWxD) |
610 x 775 x 483 mm (24 x 30.5 x 19 in)
|
622 x 768 x 483 mm (24.5 x 30.25 x 19 in)
|
775 x 768 x 483 mm (30.5 x 30.25 x 19 in)
|
762 x 686 x 401 mm (30 x 27 x 15.8 in)
|
559 x 686 x 401 mm (22 x 27 x 15.8 in)
|
Automation Package |
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. (Interfaces not available for Model 7011.)
|
沒有留言: