2014年5月1日 星期四

Fluke Hart 7015, 7009, 7008 高穩定度溫度校正槽

Fluke Hart 高穩定度溫度校正槽 Model7015, 7009, 7008
Fluke Hart Resistor Baths

溫度校正用的標準溫度源。
1.溫度產生採用三明治原理(加熱、等溫、冷卻)一體成形
2.可預先設定 8 種不同溫度在控制面板上
3.如有需要可加裝 RS-232 及 IEEE-488
4.溫度範圍從 -100 度 C 到 +550 度 C






Specifications 
7015 
7009 
7108 
Range 
0 °C to 50 °C 
0 °C to 50 °C 
20 °C to 30 °C 
Stability at 25 °C 
±0.0007 °C  (water)
±0.001 °C  (mineral oil 5011) 
±0.002 °C (water)
±0.004 °C (mineral oil 5011) 
Uniformity 
±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011) 
±0.005 °C (water)
±0.008 °C (mineral oil 5011) 
Temperature Setting 
Digital display with push-button data entry 
Set-Point Resolution 
0.002 °C; high-resolution mode, 0.00003 °C 
Display Resolution 
0.01 °C 
Digital Setting Accuracy 
±1 °C 
±0.5 °C 
Digital Setting Repeatability 
±0.01 °C 
Heaters 
500 and 1000 Watts 
Peltier heating/cooling 
Cooling Capacity 
100 to 200 Watts 
100 W in ambient 23 °C 
Access Opening 
699 x 279 mm (27.5 x 11 in) 
699 x 559 mm (27.5 x 22 in) 
356 x 356 mm (14 x 14 in) 
Bath Chamber Dimensions (HxWxD) (unobstructed space) 
699 x 279 x 330 mm
(27.5 x 11 x 13 in) 
 669 x 559 x 330 mm
(27.5 x 22 x 13 in) 
356 x 203 x 355 mm
(14 x 8 x 14 in) 
Depth 
330 mm (13 in) 
203 mm (8 in) 
Wetted Parts 
304 stainless steel 
Tank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum 
Safety Cutout 
Factory-set high temperature 
n/a 
Power 
115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify 
230 VAC (±10 %), 50 or 60 Hz, 12 A
(specify frequency) 
115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify 
Volume 
95 liters (25 gallons) 
167 liters (44 gallons) 
51 liters (13.2 gallons) 
Weight 
141 kg (310 lb.) 
150 kg (330 lb.) 
35 kg (75 lb.) 
Size (HxWxD) 
1219 x 1118 x 559 mm
(48 x 44 x 22 in) 
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in) 
489 x 413 x 559 mm
(19.25 x 22 x 25 in) 
Automation Package 
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. 
Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50 °C to protect standard resistors placed in the bath.