Fluke Hart 高穩定度溫度校正槽 Model7015, 7009, 7008
Fluke Hart Resistor Baths溫度校正用的標準溫度源。
1.溫度產生採用三明治原理(加熱、等溫、冷卻)一體成形
2.可預先設定 8 種不同溫度在控制面板上
3.如有需要可加裝 RS-232 及 IEEE-488
4.溫度範圍從 -100 度 C 到 +550 度 C
Specifications |
7015
|
7009
|
7108
|
Range |
0 °C to 50 °C†
|
0 °C to 50 °C†
|
20 °C to 30 °C
|
Stability at 25 °C |
±0.0007 °C (water)
±0.001 °C (mineral oil 5011) |
±0.002 °C (water)
±0.004 °C (mineral oil 5011) | |
Uniformity |
±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011) |
±0.005 °C (water)
±0.008 °C (mineral oil 5011) | |
Temperature Setting |
Digital display with push-button data entry
| ||
Set-Point Resolution |
0.002 °C; high-resolution mode, 0.00003 °C
| ||
Display Resolution |
0.01 °C
| ||
Digital Setting Accuracy |
±1 °C
|
±0.5 °C
| |
Digital Setting Repeatability |
±0.01 °C
| ||
Heaters |
500 and 1000 Watts
|
Peltier heating/cooling
| |
Cooling Capacity |
100 to 200 Watts
|
100 W in ambient 23 °C
| |
Access Opening |
699 x 279 mm (27.5 x 11 in)
|
699 x 559 mm (27.5 x 22 in)
|
356 x 356 mm (14 x 14 in)
|
Bath Chamber Dimensions (HxWxD) (unobstructed space) |
699 x 279 x 330 mm
(27.5 x 11 x 13 in) |
669 x 559 x 330 mm
(27.5 x 22 x 13 in) |
356 x 203 x 355 mm
(14 x 8 x 14 in) |
Depth |
330 mm (13 in)
|
203 mm (8 in)
| |
Wetted Parts |
304 stainless steel
|
Tank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum | |
Safety Cutout |
Factory-set high temperature
|
n/a
| |
Power |
115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify
|
230 VAC (±10 %), 50 or 60 Hz, 12 A
(specify frequency) |
115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify
|
Volume |
95 liters (25 gallons)
|
167 liters (44 gallons)
|
51 liters (13.2 gallons)
|
Weight |
141 kg (310 lb.)
|
150 kg (330 lb.)
|
35 kg (75 lb.)
|
Size (HxWxD) |
1219 x 1118 x 559 mm
(48 x 44 x 22 in) |
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in) |
489 x 413 x 559 mm
(19.25 x 22 x 25 in) |
Automation Package |
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE.
| ||
†Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50 °C to protect standard resistors placed in the bath. |