RUSKA 7250 / 7050 數位壓力顯示器校正器維修
Ruska 7250上蓋打開
Ruska 7050上蓋打開
Ruska sensor最重要的壓力感應器(包在恆溫槽裡)
壓力控制模組
主機板及控制板
本公司已由原廠授權維修FLUKE Ruska & DHI產品,並取得受訓結業証書
如您的Ruska產品控制壓力不穩定,精度飄移,不穩定或顯示器故障...等問題
都歡迎您與我聯絡討論,我們將提供您最快速最優惠的在台維修
特順科技有限公司 溫度壓力校正儀器設備銷售,如對儀器的使用上或是配置上有任何問題,或是維修問題,都歡迎與我們聯絡 0915-230-275 賴先生 or 03-4550133 E-mail:brightlai168@gmail.com
Specifications
| |||||
Model
| Nominal Value W | Stability, ppm 6 Months | Stability, ppm 12 Months | Max Change, ppm from 23°C ±5°C (±ppm) | Calibration Uncertainty, ±ppm |
742A-1 | 1 | 5 | 8 | 3.0 | 1.0 |
742A-10 | 10 | 5 | 8 | 3.0 | 1.0 |
742A-25 | 25 | 5 | 8 | 3.0 | 1.0 |
742A-100 | 100 | 4 | 6 | 3.0 | 1.0 |
742A-1K | 1K | 4 | 6 | 2.0 | 1.5 |
742A-10K | 10K | 2.5 | 4 | 1.5 | 1.0 |
742A-100K | 100K | 4 | 6 | 2.0 | 2.5 |
742A-1M | 1M | 6 | 8 | 2.0 | 5.0 |
742A-10M | 10M | 6 | 9 | 3.0 | 10.0 |
742A-7002 | Transit Case (holds two units) |
Specifications | |||
6020
|
6022
|
6024
| |
Range |
40 °C to 300 °C†
| ||
Uniformity |
±0.001 °C at 40 °C (water)
±0.003 °C at 100 °C (oil 5012) ±0.005 °C at 300 °C (oil 5017) | ||
Temperature Setting |
Digital display with push-button data entry
| ||
Set-Point Resolution |
0.01 °C; high-resolution mode, 0.00018 °C
| ||
Display Temperature Resolution |
0.01 °C
| ||
Digital Setting Accuracy |
±1 °C
| ||
Digital Setting Repeatability |
±0.02 °C
| ||
Heaters |
350 and 1050 watts
| ||
Access Opening (call for custom openings) |
127 x 254 mm (5 x 10 in)
|
184 x 324 mm
(7.25 x 12.75 in) | |
Depth |
305 mm (12 in)
|
464 mm (18.25 in)
|
337 mm (13.25 in)
|
Wetted Parts |
304 stainless steel
| ||
Power |
115 VAC (±10 %), 50/60 Hz, 10 A or 230 VAC (±10 %), 50/60 Hz, 5 A, specify
| ||
Volume |
27 liters (7.2 gallons)
|
42 liters (11.2 gallons)
| |
Weight |
32 kg (70 lb.)
|
36 kg (80 lb.)
| |
Size (HxWxD) |
648 x 406 x 508 mm
(25.5 x 16 x 20 in) |
813 x 406 x 508 mm
(32 x 16 x 20 in) |
699 x 483 x 584 mm
(27.5 x 19 x 23 in) |
Automation Package |
Interface-it software and RS-232 computer interface are included for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE.
| ||
†External cooling required for operation below 40 °C. Cooling coils are built into the |
Specifications | |
Range | 180 °C to 550 °C |
Stability | ±0.002 °C at 200 °C (salt) ±0.004 °C at 300 °C (salt) ±0.008 °C at 550 °C (salt) |
Uniformity | ±0.005 °C at 200 °C (salt) ±0.020 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry |
Set-Point Resolution | 0.01 °C; high-resolution mode, 0.00018 °C |
Display Temperature Resolution | 0.01 °C |
Digital Setting Accuracy | ±1 °C |
Digital Setting Repeatability | ±0.02 °C |
Heaters | 400, 1200, and 2000 Watts |
Access Opening | 127 x 254 mm (5 x 10 in) |
Depth | 305 mm (12") |
Wetted Parts | 304 stainless steel |
Power | 230 VAC (±10 %), 50/60 Hz, 10 A |
Volume | 27 liters (7.1 gal), requires 50 kg (112 lb) of bath salt |
Weight | 82 kg (180 lb.) |
Size (HxWxD) | 724 x 518 x 622 mm (28.5 x 20.4 x 24.5 in) |
Automation Package | Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |
Specifications | |
Range | 180 °C to 550 °C |
Stability | ±0.002 °C at 200 °C (salt) ±0.004 °C at 300 °C (salt) ±0.008 °C at 550 °C (salt) |
Uniformity | ±0.005 °C at 200 °C (salt) ±0.020 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry |
Set-Point Resolution | 0.01 °C; high-resolution mode, 0.00018 °C |
Display Temperature Resolution | 0.01 °C |
Digital Setting Accuracy | ±1 °C |
Digital Setting Repeatability | ±0.02 °C |
Heaters | 400, 1200, and 2000 Watts |
Access Opening | 127 x 254 mm (5 x 10 in) |
Depth | 305 mm (12") |
Wetted Parts | 304 stainless steel |
Power | 230 VAC (±10 %), 50/60 Hz, 10 A |
Volume | 27 liters (7.1 gal), requires 50 kg (112 lb) of bath salt |
Weight | 82 kg (180 lb.) |
Size (HxWxD) | 724 x 518 x 622 mm (28.5 x 20.4 x 24.5 in) |
Automation Package | Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |
Specifications | |
Range | 180 °C to 550 °C |
Stability | ±0.002 °C at 200 °C (salt) ±0.004 °C at 300 °C (salt) ±0.008 °C at 550 °C (salt) |
Uniformity | ±0.005 °C at 200 °C (salt) ±0.020 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry |
Set-Point Resolution | 0.01 °C; high-resolution mode, 0.00018 °C |
Display Temperature Resolution | 0.01 °C |
Digital Setting Accuracy | ±1 °C |
Digital Setting Repeatability | ±0.02 °C |
Heaters | 400, 1200, and 2000 Watts |
Access Opening | 127 x 254 mm (5 x 10 in) |
Depth | 305 mm (12") |
Wetted Parts | 304 stainless steel |
Power | 230 VAC (±10 %), 50/60 Hz, 10 A |
Volume | 27 liters (7.1 gal), requires 50 kg (112 lb) of bath salt |
Weight | 82 kg (180 lb.) |
Size (HxWxD) | 724 x 518 x 622 mm (28.5 x 20.4 x 24.5 in) |
Automation Package | Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |
Specifications |
7007
|
6054
|
6055
|
Range |
–5 °C to 110 °C
|
50 °C to 300 °C
|
200 °C to 550 °C
|
Stability |
±0.001 °C at 0 °C
(ethanol) ±0.003 °C at 100 °C (oil 5012) |
±0.003 °C at 100 °C
(oil 5012) ±0.005 °C at 300 °C (oil 5017) |
±0.003 °C at 200 °C
(salt) ±0.01 °C at 550 °C (salt) |
Uniformity |
±0.004 °C at 0 °C
(ethanol) ±0.007 °C at 100 °C (oil 5012) |
±0.007 °C at 100 °C (oil 5012)
±0.015 °C at 300 °C (oil 5017) |
±0.005 °C at 200 °C
(salt) ±0.010 °C at 550 °C (salt) |
Temperature Setting |
Digital display with push-button data entry
| ||
Set-Point Resolution |
0.002 °C, high res. 0.00003 °C
|
0.01 °C, high res. 0.00018 °C
| |
Display Temperature Resolution |
0.01 °C
| ||
Digital Setting Accuracy |
±1 °C
| ||
Digital Setting Repeatability |
±0.005 °C
|
±0.01 °C
| |
Heaters |
250 to 1000 W
|
250 to 1000 W
|
225 to 1800 W
|
Working Area |
178 mm dia. (7 in)
|
196 mm dia. (7.7 in)
|
107 mm dia. (4.2 in)
|
Depth |
610 mm (24 in) deep, 178 mm dia.
(7 in), removable polycarbonate cover |
610 mm deep (24 in), 196 mm dia. (7.7 in), removable SST lid
|
432 mm deep (17 in), 107 mm dia. (4.2 in), removable SST lid, special viewing channel for LIG sighting
|
Wetted Parts |
304 stainless steel
| ||
Power |
230 VAC (±10 %), 50 or 60 Hz, 14 A
(Specify frequency, contact Hart if CE mark required.) |
230 VAC (±10 %), 50/60 Hz, 10.7 A
|
230 VAC (±10 %), 50/60 Hz, 7.8 A
|
Volume |
42 liters (11.2 gal.)
|
50 liters (13.2 gal.)
|
19.8 liters
(5.2 gal., 43 kg [95 lb.] of bath salt) |
Size (DxWxH) |
470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in),
1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel |
572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel
|
572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel
|
Distance from Line of Sight to Top of Fluid |
9.5 mm (3/8 in)
|
15.9 mm (5/8 in)
| |
Automation Package |
Interface-it software and RS-232 computer interface are included for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE.
|
Specifications |
7008
|
7040
|
7037
|
7012
|
7011
|
Range |
–5 °C to 110 °C
|
–40 °C to 110 °C
|
–10 °C to 110 °C
| ||
Stability |
±0.0007 °C at 25 °C (water)
±0.001 °C at 25 °C (mineral oil) |
±0.002 °C at –40 °C (ethanol)
±0.0015 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) |
±0.0008 °C at 0 °C (ethanol)
±0.0008 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) | ||
Uniformity |
±0.003 °C at 25 °C (water)
±0.004 °C at 25 °C (mineral oil) |
±0.004 °C at –40 °C (ethanol)
±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) |
±0.003 °C at 0 °C (ethanol)
±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) | ||
Temperature Setting |
Digital display with push-button data entry
| ||||
Set-Point Resolution |
0.002 °C; high-resolution mode, 0.00003 °C
|
0.01 °C; high-resolution mode, 0.00007 °C
|
0.002 °C; high-resolution mode, 0.00003 °C
| ||
Display Resolution |
0.01 °C
| ||||
Digital Setting Accuracy |
±1 °C
| ||||
Digital Setting Repeatability |
±0.01 °C
|
±0.005 °C
| |||
Heaters |
500 and 1000 Watts
| ||||
Access Opening (call for customs) |
324 x 184 mm
(12.75 x 7.25 in) |
127 x 254 mm
(5 x 10 in) |
162 x 292 mm
(6.38 x 11.5 in) |
127 x 254 mm
(5 x 10 in) | |
Depth |
331 mm (13 in)
|
305 mm (12 in)
|
457 mm (18 in)
|
305 mm (12 in)
| |
Wetted Parts |
304 stainless steel
| ||||
Power |
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC, 50 or 60 Hz, 8 A, specify
|
115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 or 60 Hz, 9 A
(specify voltage and frequency) |
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify
| ||
Volume |
42 liters (11.2 gal)
|
27 liters (7.2 gal)
|
42 liters (11.2 gal)
|
27 liters (7.2 gal)
| |
Weight |
61 kg (135 lb)
|
63.5 kg (40 lb)
|
68 kg (150 lb)
|
56.7 kg (125 lb)
| |
Size (HxWxD) |
610 x 775 x 483 mm (24 x 30.5 x 19 in)
|
622 x 768 x 483 mm (24.5 x 30.25 x 19 in)
|
775 x 768 x 483 mm (30.5 x 30.25 x 19 in)
|
762 x 686 x 401 mm (30 x 27 x 15.8 in)
|
559 x 686 x 401 mm (22 x 27 x 15.8 in)
|
Automation Package |
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE. (Interfaces not available for Model 7011.)
|
Specifications |
7015
|
7009
|
7108
|
Range |
0 °C to 50 °C†
|
0 °C to 50 °C†
|
20 °C to 30 °C
|
Stability at 25 °C |
±0.0007 °C (water)
±0.001 °C (mineral oil 5011) |
±0.002 °C (water)
±0.004 °C (mineral oil 5011) | |
Uniformity |
±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011) |
±0.005 °C (water)
±0.008 °C (mineral oil 5011) | |
Temperature Setting |
Digital display with push-button data entry
| ||
Set-Point Resolution |
0.002 °C; high-resolution mode, 0.00003 °C
| ||
Display Resolution |
0.01 °C
| ||
Digital Setting Accuracy |
±1 °C
|
±0.5 °C
| |
Digital Setting Repeatability |
±0.01 °C
| ||
Heaters |
500 and 1000 Watts
|
Peltier heating/cooling
| |
Cooling Capacity |
100 to 200 Watts
|
100 W in ambient 23 °C
| |
Access Opening |
699 x 279 mm (27.5 x 11 in)
|
699 x 559 mm (27.5 x 22 in)
|
356 x 356 mm (14 x 14 in)
|
Bath Chamber Dimensions (HxWxD) (unobstructed space) |
699 x 279 x 330 mm
(27.5 x 11 x 13 in) |
669 x 559 x 330 mm
(27.5 x 22 x 13 in) |
356 x 203 x 355 mm
(14 x 8 x 14 in) |
Depth |
330 mm (13 in)
|
203 mm (8 in)
| |
Wetted Parts |
304 stainless steel
|
Tank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum | |
Safety Cutout |
Factory-set high temperature
|
n/a
| |
Power |
115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify
|
230 VAC (±10 %), 50 or 60 Hz, 12 A
(specify frequency) |
115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify
|
Volume |
95 liters (25 gallons)
|
167 liters (44 gallons)
|
51 liters (13.2 gallons)
|
Weight |
141 kg (310 lb.)
|
150 kg (330 lb.)
|
35 kg (75 lb.)
|
Size (HxWxD) |
1219 x 1118 x 559 mm
(48 x 44 x 22 in) |
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in) |
489 x 413 x 559 mm
(19.25 x 22 x 25 in) |
Automation Package |
Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE.
| ||
†Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50 °C to protect standard resistors placed in the bath. |