RUSKA 7250 / 7050 數位壓力顯示器校正器維修
Ruska 7250上蓋打開
Ruska 7050上蓋打開
Ruska sensor最重要的壓力感應器(包在恆溫槽裡)
壓力控制模組
主機板及控制板
本公司已由原廠授權維修FLUKE Ruska & DHI產品,並取得受訓結業証書
如您的Ruska產品控制壓力不穩定,精度飄移,不穩定或顯示器故障...等問題
都歡迎您與我聯絡討論,我們將提供您最快速最優惠的在台維修
特順科技有限公司 溫度壓力校正儀器設備銷售,如對儀器的使用上或是配置上有任何問題,或是維修問題,都歡迎與我們聯絡 0915-230-275 賴先生 or 03-4550133 E-mail:brightlai168@gmail.com
Specifications
| |||||
Model
| Nominal Value W | Stability, ppm 6 Months | Stability, ppm 12 Months | Max Change, ppm from 23°C ±5°C (±ppm) | Calibration Uncertainty, ±ppm |
742A-1 | 1 | 5 | 8 | 3.0 | 1.0 |
742A-10 | 10 | 5 | 8 | 3.0 | 1.0 |
742A-25 | 25 | 5 | 8 | 3.0 | 1.0 |
742A-100 | 100 | 4 | 6 | 3.0 | 1.0 |
742A-1K | 1K | 4 | 6 | 2.0 | 1.5 |
742A-10K | 10K | 2.5 | 4 | 1.5 | 1.0 |
742A-100K | 100K | 4 | 6 | 2.0 | 2.5 |
742A-1M | 1M | 6 | 8 | 2.0 | 5.0 |
742A-10M | 10M | 6 | 9 | 3.0 | 10.0 |
742A-7002 | Transit Case (holds two units) |
Specifications | |||
6020
|
6022
|
6024
| |
Range |
40 °C to 300 °C†
| ||
Uniformity |
±0.001 °C at 40 °C (water)
±0.003 °C at 100 °C (oil 5012) ±0.005 °C at 300 °C (oil 5017) | ||
Temperature Setting |
Digital display with push-button data entry
| ||
Set-Point Resolution |
0.01 °C; high-resolution mode, 0.00018 °C
| ||
Display Temperature Resolution |
0.01 °C
| ||
Digital Setting Accuracy |
±1 °C
| ||
Digital Setting Repeatability |
±0.02 °C
| ||
Heaters |
350 and 1050 watts
| ||
Access Opening (call for custom openings) |
127 x 254 mm (5 x 10 in)
|
184 x 324 mm
(7.25 x 12.75 in) | |
Depth |
305 mm (12 in)
|
464 mm (18.25 in)
|
337 mm (13.25 in)
|
Wetted Parts |
304 stainless steel
| ||
Power |
115 VAC (±10 %), 50/60 Hz, 10 A or 230 VAC (±10 %), 50/60 Hz, 5 A, specify
| ||
Volume |
27 liters (7.2 gallons)
|
42 liters (11.2 gallons)
| |
Weight |
32 kg (70 lb.)
|
36 kg (80 lb.)
| |
Size (HxWxD) |
648 x 406 x 508 mm
(25.5 x 16 x 20 in) |
813 x 406 x 508 mm
(32 x 16 x 20 in) |
699 x 483 x 584 mm
(27.5 x 19 x 23 in) |
Automation Package |
Interface-it software and RS-232 computer interface are included for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE.
| ||
†External cooling required for operation below 40 °C. Cooling coils are built into the |
Specifications | |
Range | 180 °C to 550 °C |
Stability | ±0.002 °C at 200 °C (salt) ±0.004 °C at 300 °C (salt) ±0.008 °C at 550 °C (salt) |
Uniformity | ±0.005 °C at 200 °C (salt) ±0.020 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry |
Set-Point Resolution | 0.01 °C; high-resolution mode, 0.00018 °C |
Display Temperature Resolution | 0.01 °C |
Digital Setting Accuracy | ±1 °C |
Digital Setting Repeatability | ±0.02 °C |
Heaters | 400, 1200, and 2000 Watts |
Access Opening | 127 x 254 mm (5 x 10 in) |
Depth | 305 mm (12") |
Wetted Parts | 304 stainless steel |
Power | 230 VAC (±10 %), 50/60 Hz, 10 A |
Volume | 27 liters (7.1 gal), requires 50 kg (112 lb) of bath salt |
Weight | 82 kg (180 lb.) |
Size (HxWxD) | 724 x 518 x 622 mm (28.5 x 20.4 x 24.5 in) |
Automation Package | Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |
Specifications | |
Range | 180 °C to 550 °C |
Stability | ±0.002 °C at 200 °C (salt) ±0.004 °C at 300 °C (salt) ±0.008 °C at 550 °C (salt) |
Uniformity | ±0.005 °C at 200 °C (salt) ±0.020 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry |
Set-Point Resolution | 0.01 °C; high-resolution mode, 0.00018 °C |
Display Temperature Resolution | 0.01 °C |
Digital Setting Accuracy | ±1 °C |
Digital Setting Repeatability | ±0.02 °C |
Heaters | 400, 1200, and 2000 Watts |
Access Opening | 127 x 254 mm (5 x 10 in) |
Depth | 305 mm (12") |
Wetted Parts | 304 stainless steel |
Power | 230 VAC (±10 %), 50/60 Hz, 10 A |
Volume | 27 liters (7.1 gal), requires 50 kg (112 lb) of bath salt |
Weight | 82 kg (180 lb.) |
Size (HxWxD) | 724 x 518 x 622 mm (28.5 x 20.4 x 24.5 in) |
Automation Package | Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |